95-8444

28

10.1

Table 14—Manual Calibration Procedure

Description

Display

Operator Action

Normal operation/no gas present Indicates the detected gas

concentration

• Ensure that the unit is set for Manual Calibration.

Refer to the “Setup Procedure.”

• If the possibility of background gases exists,

purge the sensor with clean air to ensure accurate

calibration.

Initiate calibration

Sequences through the Setup

settings

Enters the Calibration mode

-

Display shows detected gas

concentration and alternating

message:

“ZERO”

“CAL”

• Hold the Cal magnet to the CAL/RESET magnetic

switch for 7 seconds to enter the calibration mode.

(The external reset input can also be used to enter

the calibration mode if “EXT CAL YES” was selected

during the Setup procedure).

Zero calibration

Display continues showing detected

gas concentration and alternating

message:

“ZERO”

“CAL”

• When the zero reading has stabilized, hold the Cal

magnet to the CAL/RESET magnetic switch for 2

seconds to accept the displayed reading as the zero

point.

Apply calibration gas

Display shows detected gas

concentration and alternating

message:

“APLY”

“GAS”

• Apply the calibration gas to the sensor by placing

the calibration cup over the sensor and opening the

valve on the calibration gas cylinder.

Span calibration

Display shows rising gas

concentration and alternating

message:

“GAS”

“ON”

• When the reading has stabilized, hold the Cal magnet

to the CAL/RESET magnetic switch for 2 seconds

to accept the displayed reading as the span point.

The display will indicate the selected calibration gas

concentration as span.

Remove calibration gas

Display shows decreasing gas

concentration when gas is removed

and alternating message:

“CAL” “OK”

then

“RMV”

“GAS”

then

“XXXX SPAN”

• Remove the calibration gas. When the gas level

falls below the lowest alarm setpoint, the transmitter

automatically exits the Calibrate mode.

• A sensitivity reading that can be used to track sensor

life (except PointWatch) is displayed for 7 seconds

with the “SPAN” display and prior to the transmitter

returning to normal operation. Any reading over 100

indicates that the sensor is good.

• At the successful completion of the calibration, all

outputs and indicators return to normal operation.

• If faults are present, the unit will exit after the

remainder of 10 minutes. To force an exit from

calibration, make sure that the gas concentration is

below alarm setpoints and dropping, then activate

the CAL/RESET switch for 1 second.

Calibration fault indication

Display shows alternating messages

indicating the fault, then

“RMV”

“GAS”

• If a fault occurs, remove the gas and correct the fault.

After the fault has been cleared, begin calibration

again. See Table 15 - Fault Messages, Explanations

and Corrective Action.