Det-Tronics U9500 Infiniti Gas Transmitter User Manual

Page 29

Advertising
background image

95-8444

27

10.1

Routine Calibration Procedure
1. Do not remove the C7065E sensor cover.
2. Apply the calibration magnet to the U9500 calibration

target location for nine seconds, or until the U9500

display shows “zero-cal”. The initial zero baseline

output level reading will be referenced. Wait until the

U9500 display shows “aply-gas”.

3. The transmitter will now begin to record the sensor

output level in response to the ambient oxygen

concentration. If this is known to be 20.9%, an

accurate span calculation will result. If the ambient

O

2

level is less than 20.9%, it is recommended

to apply 20.9% oxygen using a Det-Tronics O

2

calibration kit.

4. When span calculations are complete, the U9500

display will indicate “cal-OK”, and then “rmv-gas”.

If using a Det-Tronics O

2

calibration kit, remove the

calibration cup and close the regulator valve. No

actions are necessary if ambient air O

2

is used.

5. The U9500 will automatically exit calibration mode

when calibration is complete. The displayed O

2

level

should be 20.9% (17.4 mA current level) if ambient

conditions are normal.

CGS SENSOR REPLACEMENT
Under certain circumstances, an existing U9500A

Infiniti gas transmitter fitted with a replacement CGS

sensor may not exit “warm up” mode after four minutes

from power up, causing a “Zero Drift” indication to be

displayed on the U9500A. This uncommon situation can

occur when a CGS pellistor resistance is slightly offset

from the previous CGS pellistor resistance. It does not

mean that the sensor or the Infiniti transmitter are bad or

out of tolerance.

There are two simple methods to solve this problem.

Preferred method:

This method requires a Hot Work Permit and

declassification of the work area.
1. Disable any alarming devices triggered by the

U9500A Infiniti transmitter alarm outputs.

2. Power up the U9500A with a new or different CGS

sensor installed.

3. Wait for the U9500A Infiniti transmitter to display

“Zero Drift” indication (approximately 4 minutes

from power up).

4. Remove the cover from the Infiniti transmitter.

5. While the Infiniti transmitter display is indicating “Zero

Drift,” simultaneously press and release the Increase

and Decrease switches on the Infiniti faceplate. The

Infiniti will exit the warm up mode and display an

incorrect gas level reading.

6. Calibrate the U9500A following the “Initial/

Replacement Sensor Calibration” procedure.

7. Replace the Infiniti J-box cover.
8. Re-enable any alarming devices triggered by the

U9500A Infiniti alarm outputs.

Alternate method:

This method can be used in a classified/hazardous area.

1. Disable any alarming devices triggered by the

U9500A Infiniti transmitter alarm outputs.

2. Power up the U9500A with a new or different CGS

sensor installed.

3. Wait for the U9500A Infiniti transmitter to display

“Zero Drift” indication (approximately 4 minutes

from power up).

4. Apply calibration gas to the sensor. After

approximately 30 seconds, the U9500A will reset,

go through the startup cycle, and display some

incorrect gas level.

5. Apply a calibration magnet and hold it in place until

the Infiniti transmitter enters the calibration mode.

The Infiniti display will read a gas indication number

alternating with “Zero Cal”.

6. Remove the calibration gas from the sensor.
7. Allow the Infiniti to zero calibrate to the new or

different sensor.

8. When prompted by the Infiniti, apply the calibration

gas to set the 50% LFL span setting.

9. Once the Infiniti accepts the new span setting, follow

the prompts to exit the calibration mode.

10. From this point, allow the sensor to fully warm up for

one hour minimum, and then calibrate the U9500A

following the normal calibration procedure in the

instruction manual.

11. Re-enable any alarming devices triggered by the

U9500A Infiniti alarm outputs.

Contact the factory if you have any questions or

comments.

Advertising