Det-Tronics U9500 Infiniti Gas Transmitter User Manual

Page 33

Advertising
background image

95-8444

31

10.1

Contaminating Gases And Vapors for Catalytic
Sensors
Gases and/or vapors that can cause permanent loss of

sensor sensitivity:

– Silicone compounds often found in oils, greases

and resins.

– Antiknock compounds such as tetra ethyl lead,

tetra methyl lead and phosphate esters.

Gases and/or vapors that can cause inhibition (temporary

loss of sensitivity) of the sensing element:

– Volatile halogenated compounds.

– Tetrachloroethylene.

– Fluorinated hydrocarbon.

– Vapors that result from hot P.V.C.

– Acids and various solvents may cause a shift in the

zero of electrochemical sensor.

– Chlorine and Chlorinated HC.

NOTE

The sensor must be periodically inspected and
calibrated following prolonged exposure to
contaminating gases and vapors.

mAINTENANCE

CaUTiOn

Be sure to secure all output devices that are
actuated by the system to prevent unwanted
activation of this equipment, and remember to
place these same output devices back into service
when the checkout is complete.

• Periodically apply a generous amount of Lubriplate

grease (part number 102868-001) to the internal

threads on the junction box cover using an

application tool. DO NOT use other lubricants, since

some materials can cause irreversible damage to the

sensing element in the gas sensor.

• To ensure reliable protection, it is important to check

and calibrate the detection system on a regularly

scheduled basis. The frequency of these checks

is determined by the requirements of the particular

installation - 90 days is typical.

• Fault detection circuitry continuously monitors for

problems that could prevent proper system response.

It does not monitor external response equipment or

the wiring to these devices. It is important that these

devices be checked initially when the system is

installed, as well as periodically during the ongoing

maintenance program.

• The system must be checked periodically in the

Normal mode to ensure that those items not checked

by the transmitter diagnostic circuitry (such as output

relays) are functioning properly.

SENSOR AND HYDROPHOBIC FILTER
REPLACEMENT

CaUTiOn

Be sure to secure all output devices that are
actuated by the system to prevent unwanted
activation of this equipment, and remember to
place these same output devices back into service
when the checkout is complete.

Catalytic Sensor Replacement
For catalytic combustible gas sensors, it is necessary to

replace the entire sensor assembly when it is defective.

Follow the procedure below to replace the sensor.

1. Remove power to the transmitter prior to replacing

the sensor.

2. Remove the transmitter cover and the wire shield

within the transmitter. If a sensor termination box

was used, remove the sensor junction box cover.

3. Unplug the sensor from the transmitter module (if a

sensor termination box was used, unplug it from the

plug inside the termination box) and unscrew it from

the conduit entry.

4. Thread the wires for the replacement sensor through

the conduit entry, then screw the sensor into the

conduit entry and plug it in. Replace the wire

shield.

5. Replace the junction box cover.
6. Re-apply power. Allow time for the unit to warm

up and stabilize (approximately 24 hours for best

results), then calibrate.

Advertising