Release the evaporator flask from the vapor tube, Insert the inlet tube – Heidolph Hei-VAP Precision User Manual

Page 21

Advertising
background image

Set-up and Start-up

Set-up and Start-up

40

41

Release the evaporator flask from the vapor tube

1. Switch off the rotation.

2. Move the evaporator flask up from the bath (see chapter “Set-up and start-up“).

3. Let the evaporator flask cool.

4. Release the vapor tube coupling clamp flask ejector nut.

WARNING

Glass breakage!

Serious injuries from shattering and breaking glass.

Check glass devices for damages (stars, breaks, etc.).

Only use perfect glass devices and replace all damaged

glassware as soon as possible.

Work carefully.

CAUTION

Unintentionally rotating drive!

Injuries to hands.

Verify that the rotation is displayed.

WARNING

Risk of injuries due to retracting or entrapment!

Risk of injury.

A risk of entanglement exists on the movable parts of the unit.

Wear suitable protective clothing for activities on the rotary

evaporator with goggles and gloves.

WARNING

Hot media!

Burns.

Do not touch the interior and the upper edge of the heating bath, the

evaporator flask and the heating bath liquid.

Wear suitable heat protection gloves when changing the evaporator

flask.

5. Release the flask from the vapor tube (Figure 4-6 (7)) by turning the vapor tube

coupling clamp flask ejector nut counter-clockwise by supporting the flask.

6. Hold the evaporator flask and release the flask clamp.

7. Carefully remove the flask.

Insert the inlet tube

The inlet tube consists of:

Stopcock (Stopper)

PTFE tube

Drain-off plate

WARNING

Glass breakage!

Serious injuries from shattering and breaking glass.

Check glass devices for damages (stars, breaks, etc.).

Work carefully.

1

PTFE tube

3

Stopcock

2

Drain-off plate

Figure 4-15: Insert the inlet tube

3

2

1

1. Shorten the PTFE tube (1) to the desired length.

2. Grease the stopcock (3) (optional) for ideal vacuum.

3. Position the drain-off plate (2) so that the condensed liquid cannot return to the

evaporator flask via the PTFE tube (1).

4. Insert the inlet tube

Advertising