Operation

Operation

82

83

Online Help

Shows Online-Help instead of the actual values in the menu

SET

pressure

, AUTO

easy

, AUTO

accurate

, Gradient, Favorites and

Setup.

Safety pressure

If the system reaches the setting of the Safety Pressure

point, the vacuum control and flask rotation is stopped, the

evaporator is vented and the motor lift is raised (ML only).

Setting range is between 900 and 1200 mbar.

Heating Bath Media

Allows switching between different heating bath media and

sets a limit to the maximum heating bath temperature for

the media for operator safety.

Max. Temp. Heating

Bath

Sets the limit of the maximum heating bath temperature

which can be set. The chosen bath media has an influence on

the temperature range. Maximum set range between 20 and

210 °C.

Language

Sets language of the interface.

Bath Power at End

Sets the state of the bath after the process has ended. When

setting is “off” the bath is powered down after the process is

terminated.

Brightness Display

Changes brightness of the display.

Calibrate p-sensor

Opens sub-menu for calibration of the pressure sensor.

Error history

Shows the errors and history of occurrence for service data.

Reboot setup

All settings of the Setup are put into the state of factory

setting when Reboot engaged.

Scale time axis

Sets range of the visible time frame in the graphical “display

actual values”. Set range is between 1 and 120 minutes.

Setup

Menu topology

Actual value display for all modes

Menu entry

Value range

Vacuum

1 and 1200 mbar; default value: 1000 mbar

Vapor

Vapor temperature measured at the vapor temperature

sensor.

Rotation

10 and 290 rpm in 1 rpm increments; default value: 100 rpm

Heating bath

20 °C and the max. temperature of the heating bath (to be

assigned in Setup), however, no more than 210 °C.

Softkey 1

Start Vacuum

Softkey 2

Start all

Stop all

Softkey 3

Max. Vacuum

Softkey 4

Stop Vacuum

SET

pressure

Mode

Menu entry

Value range

Heating bath

20 °C and the max. temperature of the heating bath (to be

assigned in Setup), however, no more than 210 °C.

Rotation

10 and 290 rpm in 1 rpm increments; default value: 100 rpm

Vacuum

1 and 1200 mbar; default value: 1000 mbar

Hysteresis

1 to 50 mbar; default value: 10 mbar

Timer

1 to 1440 min; in 1 min; default value: 0 min; 0=off

Accept (OK)

Softkey 1

Store Favorite

Softkey 4

Esc