General safety instructions, Intended use, Improper use – Heidolph Hei-VAP Precision User Manual

Page 6: Use in explosion-prone areas, Responsibilities of the operator

Advertising
background image

10

11

Basic safety instructions

Basic safety instructions

The rotary evaporator is constructed according to the current state of the art and

accepted safety regulations. However, risks still exist during the installation, operation

and maintenance of the product.

Observe safety instructions and warnings.

The basic safety instructions in this chapter are supplemented in the following chapters

of the operating instructions by concrete warnings. These warnings will precisely explain

how you must conduct yourself to protect yourself, other persons and objects from

injuries or damages.

These instructions are for the following evaporator models, the Hei-VAP Precision rotary

evaporator.

Always keep the instructions available.

Pass the instructions on to subsequent owners.

General Safety Instructions

The rotary evaporator may only be used

in an operational condition consistent with full functionality of equipment,

for the intended use described in section 2.2,

if the operator has the required safety protection and awareness to hazards,

if the instructions of these operating instructions are observed.

Malfunctions, especially those that may affect safety, must be repaired immediately!

Intended use

Hei-VAP Value, Hei-VAP Advantage, Hei-VAP Precision are rotary evaporators for:

distillation or evaporation of solvents

purification of chemicals, substances, mixtures and preparations

processing reaction batches

drying of powder

Improper Use

Applying excess pressure to the equipment is not permitted and can result in

explosion of the system.

Do not use the equipment in explosion-prone areas based on local ordinance and

compliance of general laboratory equipment. The equipment is not protected against

explosions in excessively volatile and poorly ventilated environments.

Use in explosion-prone areas

Do not use the equipment in explosion-prone areas. The equipment is not protected

against explosions. It is not equipped with Ex or ATEX protection.

Responsibilities of the operator

Only operate the equipment in perfect condition, contact a certified Heidolph Service

representative if equipment does not meet the outlined conditions for operation.

Verify that only qualified personnel operate the equipment with the appropriate per-

sonal protective protection.

Verify that personnel have received safety instructions for working responsibly and

safely in the laboratory.

Verify that the rotary evaporator is set up in a safe and suitable location.

Verify that the basic device is only plugged to or unplugged from the heating bath if

the main switch is switched off and/or the main power supply is disconnected.

Depending on the media (chemicals) used:

Verify that the set up and the operation are only within the designated buildings with

the proper equipment for laboratories.

Verify that the rotary evaporator is only operated in conjunction with a ventilation

system (refer to DIN EN 14175 and DIN 12924).

Ventilation system:

- at least 10 times the air exchange

- monitored for malfunctions

Any other or additional use is considered not to be in accordance with its designated

use.

The manufacturer is not liable for damage resulting from this action.

The operator alone carries the responsibility to comply with the intended use outlined

above. Observing these instructions and all references, especially safety instructions,

as well as the adherence to inspection and maintenance requirements (see chapter

“Maintenance, cleaning, service” – “Maintenance”) are a part of the designated use of

Hei-VAP rotary evaporator.