Replacing the fuses, Readjust the vacuum seal (glassware set g6) – Heidolph Hei-VAP Precision User Manual

Page 49

Advertising
background image

Maintenance, cleaning, service

96

97

Maintenance, cleaning, service

Replacing the fuses

The fuses are located on the bottom of the unit.

Readjust the vacuum seal (glassware set G6)

1

Valve

2

Center piece G6

Figure 8-2: Readjust the vacuum seal - glassware set G6

The seal in the valve (1) can be readjusted with a screwdriver.

1. Screw the valve into the G6 center piece (2) up to the stop.

2. Turn the adjusting screw clockwise until a white ring can be seen at the shaft

of the glass.

WARNING

Glass breakage!

Cuts.

Work carefully; do not readjust the adjusting screw too much.

1

2

Figure 8-3: Fuse on the bottom of the unit

CAUTION

The drive unit and the heating bath may be switched on

accidentally!

Injuring and burning hands.

Verify that the power switch for the base unit and the heating bath is

switched off via the pushbutton.

WARNING

Risk of injuries due to retracting or entrapment!

Risk of injury.

A risk of entanglement exists on the movable parts of the unit.

Wear suitable protective clothing for activities on the rotary

evaporator with goggles and gloves.


1. Disconnect the unit from the power plug.

2. Remove evaporating flask.

3. Cool the heating bath liquid and remove the heating bath from the base

unit by disconnecting the power plug.

4. Carefully tilt the unit and pour out the bath media.

5. Dry the bath and turn it over with the bottom facing upward.

6. Remove the defective fuse by using a screwdriver.

7. Insert a suitable new fuse and tighten with a screwdriver.

The fuse has been replaced. The unit can be returned to active heating operation.