Errors and Troubleshooting

Errors and Troubleshooting

88

89

Error / message

Cause

Remedy

Motor lift is not

functioning

Master switch is deactivated

Activate main switch

Master switch is defective

Contact service

Fuses defective or blown fuse

Replace fuses (see

chapter “Maintenance,

cleaning, service” –

“Maintenance”)

Lift is at end stop

Activate other arrow key

Mechanical system / motor is

defective

Contact service

Height stop is not set

accurately

Perform height adjust-

ment (see chapter

“Set-up and start-up“ –

“Glassware sets)

No evacuation

Power switch Rotavac vario

control or Rotavac vario tec is

not activated.

Vacuum valve is defective.

Activate the power switch

Fuses defective or blown fuse

Replace fuses

Insufficient vacuum

System leaks

Check seals, tubing and

connections

Check stopcock, grease if

necessary

Vacuum pump is defective

Observe manufacturer‘s

instructions of the vacu-

um pump

Unit suddenly

switches off (Hei-VAP

Precision and Hei-VAP

Advantage)

Timer is programmed

Check and switch off the

timer, if necessary (see

chapter “Operation” –

“Timing functions”)

Fuses defective or blown fuse

Replace fuses (see

chapter “Maintenance,

cleaning, service” –

“Maintenance”)

Additional conditions

Tabelle 7-1: Error codes

Additional conditions

Reaction

Auto

accurate

sensor is not

connected

‒ Change to the SET

pressure

mode

‒ The current pressure value is accepted as target value,

similar to Hold/Accept in the SET

pressure

mode

‒ Auto

accurate

mode cannot be selected

Vapor sensor is

defective

‒ The actual value curve “- - -“ is displayed instead of the

vapor temperature.

The vapor sensor is not

connected

‒ The vapor temperature is not displayed in the actual

value display

Heater can only be

activated at > 0 °C

Table 7-2: Additional conditions

Error / message

Cause

Remedy

Unit suddenly switches

off (Hei-VAP Precision)

When pressure is detected,

the Hei-VAP Precision is

emergency stopped.

Check the settings in the

setup menu.

The Hei-VAP Precision is

reset to the starting

condition when pressing

any key.

Calibrate the vacuum

sensor (see chapter

“Operation” – “More

Functions”)