Menu display, Auto, Mode – Heidolph Hei-VAP Precision User Manual

Page 36: Actual value display, Softkey “start all” and “stop all

Advertising
background image

Operation

Operation

70

71

Menu display

The following values can be changed in the menu view:

Selecting the entry by turning the Hei-GUIDE (6):

1. Call up the selected entry to be changed by pressing the Hei-GUIDE (6)

(entry will be highlighted in orange).

2. Set the value by using the rotary pushbutton controller Hei-GUIDE (6).

3. Pressing it again confirms the change. Softkey “Esc“ (1) terminates the change

of the currently processed entry.

4. The new settings are accepted by the “accept” entry. It is changed to the actual

value display.

All changes in the menu view are discarded with softkey “Esc“ (1).

The parameters are stored under a name to be selected by using softkey

“Store Favorite“ (1).

Heating bath

Setting of the bath temperature. Setting range is between

20 °C and maximum heating bath temperature (MAX TEMP is

adjustable in Setup) with a highest range of 210 °C.

Rotation

Setting of rotation speed of the distillation flask. Setting

range is between 10 and 280 rpm.

Timer

Setting of the timer to shut down the process. Setting range

is between 1 and 1440 minutes. For continuous operation

choose “off” for Timer setting. The heating bath at the end

of a timed run can be set to turn off (Parameter found in

Setup).

Pump Type

Selects the connected pump type. The internal control

parameters are adjusted according to the connected pump

(only by using rpm controlled pump Rotavac vario control or

Rotavac vario tec.)

Auto

easy

mode

The required vacuum is automatically calculated based on a pressure measurement in

the “AUTO

easy

” mode. The boiling pressure is detected and maintained automatically.

The “AUTO

easy

” mode is configured for easy handling and for common heating bath and

cooling media temperature. In case the amount of steam produced is insufficient and the

evaporation does not start, increase the heating bath temperature. Decrease the cooling

media temperature or heating bath temperature in case of insufficient reclaim of solvent.

Actual value display

Prerequisite: The evaporator is in the AUTO

easy

mode

Rotation speed and heating bath can be changed as described in chapter “Operating the

control panel”

Softkey “Start all” and “Stop all”

A distillation can be started and stopped automatically by using softkey “Start all” (1) and

“Stop all” (1), as described in chapter “Operation” – “Softkeys”.

Auto

accurate

Mode

The vacuum in the Auto

accurate

mode is controlled by the temperature sensor Auto

accurate

.

This mode offers the following advantages:

‒ Only low quantities of solvent are suctioned by this vacuum pump.

‒ The vapor point is detected automatically and supplemented for mixtures.

‒ Multiple solvent distillation is possible without any manual adjustments.

‒ Knowledge of the thermodynamic material data is not required.

The system controls the vacuum so that a temperature increase occurs on the AUTO

accurate

sensor via the produced vapor quantity. The temperature increase refers to the

starting temperature during the de-activated process. It is also necessary that a constant

temperature (Heidolph Rotacool or Rotachill recommended) is present in the condenser

prior to the start of the process. The Hei-VAP Precision automatically checks the

temperature consistency and informs the user, if necessary. The temperature difference

between the condenser temperature and the temperature at the AUTO

accurate

sensor used

for controlling the vacuum will be increased when increasing the Distillation rate (dT).

This results in increasing the distilled solvent quantity per time.

Temperature sensor Auto

accurate

sensor (569-00040-00) is installed (see chapter

“Accessories, spare parts” – “Accessories”). The cooling medium has run for 15 min

(Auto

accurate

sensor shows a constant temperature).

A distillation with optimal recurring data results at low values for the

Distillation rate (dT) (1-3).

Low values result in a slow distillation and high values to a quick

distillation.

Select the optimal general conditions.

Auto

accurate

mode can only be run in G3 and G6 glassware

configurations.