Operation

Operation

78

79

Start pressure ramp

The pressure ramp is started with softkey “start all“ (1) or “start vacuum“ (1).

The pressure/time segments are processed.

Note: The maximum attainable increase of the pressure/time curve is limited by the

pump capacity and the leak rate of the system.

The process is terminated when reaching the last entry. The vacuum control is switched

off, the system is ventilated. The lift is raised with a motorized lift. The heating bath is

switched off depending on the setting in the setup.

Calling up the pressure ramp

Pressure ramps can be called up via the Favorites menu.

The programmed pressure ramps can be edited when the process is de-activated by

changing the respective parameters (see chapter “Operation” – “Gradient function”).

Softkey Graph

The pressure curve is displayed in the graphic view of the actual value display. The time

is illustrated in [min] on the x axis and the associated pressure in [mbar] on the y-axis.

1. Select the pressure/time step to be changed by turning the rotary pushbutton

controller Hei-GUIDE (6).

2. Confirm the selection by pressing the Hei-GUIDE (6).

3. The value can be changed by turning the Hei-GUIDE (6) to the left or right.

4. The current setting is accepted by pressing the Hei-GUIDE (6).

The data record can be saved by using softkey “Store Favorite“ (1).

Softkey “Esc“ (1) discards the changes and returns to the actual value display.

A graphic overview of the programmed ramp is displayed by using softkey “Graph“ (1).

The gradient diagram presents the time on the x-axis in [min] and the related pressure

on the y-axis in [mbar].

The scaling of the x and y axes is illustrated so that all of the entered times or pressures

are illustrated.

The entry “enter basic settings” returns to the time-unrelated basic values of the heating

bath, rotation, hysteresis, and the max. power pump.

Softkeys

Softkey “Start all” and “Stop all”

A distillation can be started and stopped automatically by using softkey “Start all” (1)

and “Stop all” (1).

The unit is operational.

The immersion depth of the evaporator flask is set (see chapter “Set-up and start-up“

– “Glassware sets”).

The desired settings are completed.

Caution - Glass breakage! The immersion depth of the evaporator flasks

must be set correctly to prevent lowering of flask into bottom of heat

bath on motor lift models.

Switching over between “graph” and “values” is not possible when the

vacuum control unit is switched off.

Only the section for which pressure values are available is shown on the pressure axis.

The displayed time range on the x-axis is set in the setup. If the set time range is

exceeded, the time window will be continuously shifted.

”Start all”

The lift moves down (only for the motor lift unit), rotation, vacuum control and timing are

started.

All parameters of the distillation can be changed during the course of

the “start all” function.

”Stop all”

The lift moves upward (only for the motor lift unit), rotation, vacuum control and timing

are stopped.