Errors and Troubleshooting

87

Operation

86

Favorites

Menu entry

Value range / status

(empty)
[name of Favorite]

Apply

Rename

Copy

Delete

Back

Softkey 4

Esc

Cancelling the entry and backing up on step, no storage of

changes

Menu entry

Value range / status

Online-Help

On/Off

Safety pressure

900 … 1200 mbar

Heating bath media

Water; PEG; OIL; Default value: Water

Max. Temp. Heating

Bath

20 to 210 °C [max. heating bath temp.] in 0.5 °C; default

value: 20 °C

Water: = 90 °C; 20 ... 95 in 1 °C

PEG, OIL = 210 °C; 20 ... 210 °C in 1 °C

Language

German, English, Spanish, French, Italian

Bath Power at End

On/Off; default value: On

Brightness Display

yes/no; default value: No

Calibration of the

p-sensor

yes/no; default value: No

Error history

yes/no; default value: No

Reboot setup

yes/no; default value: No

Scale time axis

Setting range of 1 to 120 minutes.

Accept (OK)

Setup

Softkey 4

Esc

Cancelling the entry and backing up on step, no storage of

changes

General errors

Error / message

Cause

Remedy

Unit cannot be

switched on

Main plug not connected to

the main power supply

Connect the main plug

with the power supply

Fuses defective or blown fuse

Replace fuses

(see chapter

“Maintenance,

cleaning, service” –

“Maintenance”)

No heating function

Master switch is deactivated

Activate main switch

Fuses defective or blown fuse

Replace fuses

(see chapter

“Maintenance,

cleaning, service” –

“Maintenance”)

Master switch is defective

Contact service

The connecting cable of

the heating bath is not

connected.

Connect the connecting

cable of the heating bath

The heater of the heating

bath is defective

Contact service

The maximum temperature

limiter responded

If the medium was in

the heating bath: Let the

heating bath cool and

reset the maximum

temperature limiter

If the medium was not

in the heating bath:

Contact service

Drive unit is not

rotating

Master switch is defective

Contact service

Speed controller is on the left

stop (Hei-VAP value)

Turn the speed controller

to the right

Drive unit is defective

Contact service

Fuses defective or blown fuse

Replace fuses

(see chapter

“Maintenance,

cleaning, service” –

“Maintenance”)