Operation

Operation

68

69

WARNING

Imploding glass devices!!

Serious injuries from shattering glass.

Check glass devices for damages (stars, breaks, etc.).

Only use perfect glass devices and replace all damaged glassware

as soon as possible.

SET

pressure

mode

The vacuum is controlled and maintained at the set pressure in the SET

pressure

mode.

Menu display

The following values can be changed in the menu view:

Heating bath

Setting of the bath temperature. Setting range is between

20 °C and maximum heating bath temperature (MAX TEMP is

adjustable in Setup) with a highest range of 210 °C.

Rotation

Setting of rotation speed of the distillation flask. Setting

range is between 10 and 280 rpm.

Vacuum

Setting of the absolute pressure in the rotary evaporating

system. Setting range is between 1 and safety pressure

(SAFETY PRESSURE is adjustable in Setup) with a high limit

of 1200 mbar.

Max. power pump

Vacuum pump (Vario Models) throughput % to maximize

pump power. Setting range is between 10 and 100 %.

Reduction of the throughput leads to higher solvent

recovery rates.

Hysteresis

The value of the hysteresis indicates how much the pressure

may rise (in mbar) before the vacuum valve is opened.

Setting range is between 1 and 50 mbar (only by using valve

controlled pumps Rotavac valve control, Rotavac valve tec, or

House vacuum source).

Timer

Setting of the timer to shut down the process. Setting range

is between 1 and 1440 minutes. For continuous operation

choose “off” for Timer setting. The heating bath at the end

of a timed run can be set to turn off (Parameter found in

Setup).

Selecting the entry by turning the Hei-GUIDE (6)

1. Call up the selected entry to be changed by pressing the Hei-GUIDE (6) (entry will be

highlighted in orange).

2. Set the value by using the rotary pushbutton controller Hei-GUIDE (6).

3. Pressing it again confirms the change. Softkey “Esc“ (1) terminates the change of the

currently processed entry.

4. The new settings are accepted by the “accept” entry. It is changed to the actual value

display.

All changes in the menu view are discarded with softkey “Esc“ (1).

The parameters are stored under a name to be selected by using softkey “Store

Favorite“ (1).

Actual value display

The evaporator is in the SET

pressure

mode

Rotation speed and heating bath can be changed as described in chapter “Operation“ –

“Setting the rotation speed” and “Setting the heating bath temperature”.

Vacuum control

1. Select the entry “vacuum” by turning the Hei-GUIDE (6) and calling up the selected

entry to be changed by pressing the Hei-GUIDE (6) again.

2. Set the absolute pressure by using the rotary pushbutton controller Hei-GUIDE (6).

The pressure is readjusted synchronously at an active vacuum control.

3. Pressing it again confirms the change.

4. The entry is exited after 10 seconds, in order to rule out accidental adjustments.

The setting range of the menu entry “vacuum” can be set between 1

and the safe pressure in [mbar], however, not higher than 1200 mbar.

Softkey “Start all” and “Stop all”

A distillation can be started and stopped automatically by using softkey “Start all” (1)

and “Stop all” (1), as described in chapter “Operation“ – “Setting the rotation speed” and

“Setting the heating bath temperature”.