Gradient function, Programming the pressure ramp (menu display) – Heidolph Hei-VAP Precision User Manual

Page 39

Advertising
background image

Operation

Operation

76

77

Copy data record

The data record is copied and saved under a new name. An entry of the new program

name will be required. A duplication of the issuance of the program name will be

checked.

Delete data record

Deletes the data record permanently.

Gradient function

A pressure ramp can be programmed with the gradient function, with which various

pressure values can be accessed. The respective pressures are controlled linear over

time. A pressure drop, i.e. by 50 mbar within 10 minutes, therefore results in a slight

increase of the pressure/time curve than the same pressure reduction in 5 minutes. The

maximum attainable increase of the pressure/time curve is limited by the pump capacity

and the leak rate of the system.

Rename the data record

The name can be entered similar to the “save” function. The system checks for

duplicates when naming.

Programming the pressure ramp (menu display)

The user can set the pressure and time in each line with gradient increments (but not the

sum time (total), since this is calculated automatically). The selected element is always

highlighted. The displayed gradient increments and the special elements on the right of

the menu entries are automatically scrolled upward and downward, since there is a total

of 20 gradient increments.

The currently selected entry (a pressure or a time) is called up by pressing the

Hei-GUIDE and the user can set the desired value by turning the Hei-GUIDE. Pressing the

Hei-GUIDE again will confirm the set value.

Entering and confirming value = “---” at “Vacuum” or “Duration” will delete the respective

values. The following value pairs slide upward by one increment.

Last increment with a value of > 0 at duration results in the total processing period.

1. The data record can be saved by using softkey “Store Favorite“ (1).

2. Softkey “Esc“ (1) discards the changes and returns to the actual value display.

3. The gradients can be entered in sub-menu “Enter pressure/time steps“.

4. Up to 20 gradient steps can be entered, whereby each gradient step consists of

pressure and time.

Heating bath

Setting of the bath temperature. Setting range is between

20 °C and maximum heating bath temperature (MAX TEMP is

adjustable in Setup) with a highest range of 210 °C.

Rotation

Setting of rotation speed of the distillation flask. Setting

range is between 10 and 280 rpm.

Max. power pump

Setting the maximum suction power of vacuum pump.

Setting range from 10 to 100%. A reduction of the suction

power results in a higher return rate of the solvent.

Hysteresis

Setting the pressure difference of the target value at which

the vacuum is re-adjusted. Setting range 1 – 50mbar

(only by using value controlled pump Rotavac valve control

or Rotavac valve tec, or House vacuum.)

Enter Pressure/Time

Steps

Changes the sub-menu for programming the pressure-time

ramps.

The pressure and time can be set in each line. However, the Sum of time

cannot be set, since this is automatically calculated.

The starting point is always at a time value of 0 and the atmospheric

pressure.

The following values can be changed in the menu view: