Errors and Troubleshooting

Errors and Troubleshooting

90

91

Hei-VAP Precision

Error / message
error text

Cause

Remedy

The temperature sensor
for calculating the vapor
temperature is not
connected, defective or
the temperature is above
the measured range.

‒ Vapor sensor broken
‒ not connected
‒ above the temp. range

Let the boiling temperature

sensor cool

Connect the vapor

temperature sensor

Perform visual check
Contact service

The vapor temperature
sensor is defective or the
temperature is below the
measured range

‒ Short circuit of the vapor

sensor

‒ below the temp. range

Heat the boiling

temperature sensor

Perform visual check
Contact service

The Auto

accurate

sensor

is defective or the
temperature is below the
measured range.

Auto

accurate

sensor short circui

t

‒ below the temp. range

Heat the Auto

accurate

sensor

Perform visual check
Contact service

The Auto

accurate

sensor

is defective or the
temperature is above the
measured range.

Auto

accurate

sensor is broken

‒ not connected
‒ above the temp. range

Let the Auto

accurate

sensor cool

Connect the Auto

accurate

sensor

Perform visual check
Contact service

The transportation
protection device is active,
please raise the lift and
briefly switch the unit off
and then on again.

‒ The transportation protection

device is still activated.

Please raise the lift and

briefly switch the unit off and
then on again.

An error occurred in the
data bus.

‒ Cable connection between

the control panel and the
base unit was disconnected.
An error occurred in the data
bus.

Restore the cable

connection between the
control panel and the base
unit. Switch the power
switch off and on again!

Contact service

The overheating
protection was triggered
or the heater is defective.

‒ Heater is defective, i.e. the

overheating protection was
triggered.

If the medium was in the

heating bath: Let the heating
bath cool and reset the
maximum tem-perature
limiter

If the medium was not in

the heating bath: Contact
service

Error / message
error text

Cause

Remedy

A temperature above the
safety temperature was
measured in the heating
bath. The temperature
measurement is defective
or the medium filled in
was too hot.

‒ Heating bath safety

temperature is exceeded

Switch off units
Cool the heating bath

medium

Contact Service if media was

not filled in

The heating bath is
completely evaporated.
Please refill the heating
bath medium.

‒ Heating bath is completely

evaporated.

Deactivate the unit
Check if the maximum

temperature limiter has
responded. If yes, reset
(see chapter “Maintenance,
cleaning, service” –
“Maintenance”)

Fill in heat transfer fluid

An illogical temperature
difference was measured
between the heating bath
sensors. The temperature
measurement is defective.

‒ Heating bath sensor variance

> 10 K

Contact service

The temperature sensor
for the heating bath
control is defective.
(hardware monitoring
unit)

‒ Sensor monitoring via

hardware

‒ Heating bath sensor is broken

or short circuited

Contact service

The temperature sensor
for the heating bath
control unit is broken or
bath medium that is too
hot was filled in.

‒ Heating bath sensor is broken
‒ Heating bath is more than

217 °C

Contact service

The temperature sensor
for the heating bath
control unit is defective or
bath medium that is too
cold was filled in.

‒ Heating bath sensor short

circuit

‒ Heating bath is below 0 °C

Switch off units
Cool the heating bath

medium

Contact Service if media was

not filled in

The lift motor is defective
or blocked.

‒ Lift end stop switch is not

reached after 10 seconds.

Contact service