Maintenance, cleaning, service

Maintenance, cleaning, service

Maintenance

Only use genuine parts approved by the manufacturer!

The vapor tube and PTFE 26 vacuum seal must be routinely serviced.

1. Remove and clean the vapor tube and the PTFE 26 vacuum seal (see chapter “Set-up

and start-up“ – “Glassware sets”).

2. Check the PTFE 26 vacuum seal for damages and wear, replace if necessary.

Resetting the maximum temperature limiter

In the event that the temperature controller malfunctions and the heating bath

temperature exceeds the maximum temperature of 250 °C, the maximum temperature

limiter will switch off the heating bath.

The heating bath is cooled and disconnect from power supply. Power-switch off.

1. Remove the heating bath liquid from the heating bath.

WARNING

Hot surfaces!

Burns.

Cool the heating bath.

WARNING

Verify that the basic device is only plugged to or unplugged from the

heating bath if the main switch is switched off and/or the main power

supply is disconnected.

Figure 8-1: Maximum temperature limiter

1

2. Press the maximum temperature limiter (1), possibly with a pointed object.

Evacuate the heating bath

1. The heating bath is cooled and disconnect from power supply. Power-switch off.

2. Take the heating bath from off the device.

WARNING

Verify that the basic device is only plugged to or unplugged from the

heating bath if the main switch is switched off and/or the main power

supply is disconnected.

3. The edge of the heating bath is designed to easily evacuate the heating bath.

Evacuate the heating bath.