Connect the base unit vacuum box – Heidolph Hei-VAP Precision User Manual

Page 28

Advertising
background image

Set-up and Start-up

Set-up and Start-up

54

55

1

USB PC connection

6

Ventilation / inert gas

2

Vacuum pump/Vacuum valve

connection

7

Hei-VAP connection

3

Suction side of vacuum

8

Switch box connection

4

Connecting line outlet to the heating

bath

9

Vacuum sensor connection

5

Vacuum box connection

Figure 4-32: Vacuum box (for Hei-VAP Precision)

1. Place the vacuum box on the guiderails behind the heating bath. The vacuum box

bracks must be located above the holes.

2. Place the enclosed screws (2 pieces) in the holes with thread through the bracket of

the vacuum box.

3. Tighten the fastening screw.

4. Insert the necessary or applicable connections (2, 3, 4, 5, 7, 8) into the Vacuum box.

1

2

3

4

9
8

7

6

5

CAUTION

Supply voltage and the voltage information on model tag do not

match!

Damage due to incorrect voltage.

Verify that the voltage specified on the model tag corresponds with

the mains voltage.

Only disconnect the basic device with the heating bath if the device is

de-energized. Unplug main power supply.

CAUTION

The drive unit and the heating bath may be switched on

accidentally!

Injuring and burning hands.

Risk of entanglement

Injuries to hands.

Verify that the power switch for the base unit and the heating bath

is switched off by the rocker switch.

Country-specific equipment connection cable

The design of the equipment is equipped with a Euro connector (EN

50075) for 230/240 V by default, and with a US standard plug (NEMA

Pub. No. WDI. 1961 ASA C 73.1 for a design for 120 V. 1961, page 8,

15 A 125 V). The main power cable of the device has an integrated

protective ground connection.

Please note if the appropriate equipment connection cable for your

country was included and use it.

Equipment connection cable for:

‒ EU

‒ Great Britain

‒ Switzerland

‒ USA

In order to operate the equipment in a country with a different

connector system, one can use approved adapters only or have a

licensed electrician replace the plug with one suitable and approved for

the power supply with protective ground connection.

The plug sockets are located on the back of the unit.

The unit is switched off.

The heating bath is installed on the base console.

Connect the base unit

Vacuum box

Advertising