Errors and Troubleshooting

Errors and Troubleshooting

92

93

Error / message
error text

Cause

Remedy

The set safety pressure
was exceeded.

‒ Pressure sensor safety

pressure is exceeded.

Deactivate the unit
Check if the pressure sensor

has responded. If yes, reset
(see chapter “Maintenance,
cleaning, service” –
“Maintenance”)

Contact service

The vacuum box is not
connected or defective.

‒ Vacuum box is not connected

or defective

Check the connection

between the vacuum box
and the evaporator.

Contact service

The aeration valve is
defective.

‒ Vacuum valve short circuit

Contact service

The vacuum valve is not
connected or defective.

‒ The vacuum valve is not

connected

‒ The connection between the

vacuum box and the vacuum
valve is interrupted

Connect the vacuum valve
Check the connection

between the vacuum box
and the vacuum valve.

The vacuum valve is
defective.

‒ The vacuum valve is short

circuited

Replace the vacuum valve

The vacuum pump is not
connected or defective.

‒ The vacuum pump is not

connected

‒ The connection between the

vacuum box and the vacuum
pump is interrupted

Connect the vacuum pump
Check the connection

between the vacuum box
and the vacuum pump.

The rotation drive was
overloaded or is defective.

‒ Drive unit is defective
‒ Motor overload shutdown

Deactivate the unit
Stop drive unit for cooling

down

Contact service

Table 7-3: Troubleshooting table - Hei-VAP Precision

The rotary evaporator is checked after “power on” in an initialization phase for connected

components. The function scope of the rotary evaporator is based on the connected

units.

Missing component

Description and response of
the evaporator

Remedy

Vacuum box

No vacuum box is connected:
‒ Actual value display of the

vacuum is not displayed

‒ The vacuum entry is not
displayed
‒ “Auto

easy

“, “Auto

accurate

“ and

“Gradient“ modes are not

displayed.

Note: An evaporator can be
operated with a separate pump
stand in this condition.

Connect component
Reset via power off/on

Vacuum valve or
vacuum pump

No vacuum valve or vacuum
pump is connected.
‒ Vacuumcontroller is not
active
‒ “Auto

easy

“, “Auto

accurate

“ and

“Gradient“ modes are not

displayed.

Note: An evaporator can be
operated with a separate pump
stand in this condition.

Connect component
Reset via power off/on

Temperature sensor
T boiling

The temperature sensor for
calculating the boiling
temperature is not connected:
‒ The vapor temperature is not

displayed

Connect component

Temperature sensor
Auto

accurate

sensor

The temperature sensor for
controlling the Auto

accurate

mode is

not connected:
‒ “Auto

accurate

“ mode is not

displayed

Connect component

Table 7-4: Missing Components