INFICON HLD6000 Refrigerant Leak Detector Interface Description User Manual

Page 12

Advertising
background image

12

ASCII Protocol

*STATus:CAL

status of calibration:
READY
WAIT_STABLE_INT_CAL
PHASE_INT_CAL
FACT_INT_CAL
KEEP_PROBE_STILL_INT_CAL
WAIT_STABLE_PROOF
ACK_NACK_PROOF
KEEP_PROBE_STILL_PROOF
WAIT_STABLE_EXT_CAL
PHASE_EXT_CAL
FACT_EXT_CAL
EEP_PROBE_STILL_EXT_CAL
PHASE_CAL_CO2
LEAK_CAL_CO2
PROBE_NOT_STILL_CAL_CO2
AIR_START_CAL_CO2
AIR_CAL_CO2
PROBE_NOT_STILL_AIR_CAL_CO2
RESULT_OK
RESULT_TOO_HIGH
RESULT_TOO_LOW
PROOF_OK
PROOF_NOT_OK
PROOF_IMPOSSIBLE
EXTERN_NOT_STABLE
TIMEOUT
RESULT_OFFSET_TOO_HIGH
RESULT_DIFFERENCE_TOO_SMALL

260

R

*STATus:CALHist

Calibration history
*STATus:CALHist Last cal history entry
*STATus:CALHist:1 Entry 1 (newest)
*STATus:CALHist:2 Entry 2
...
*STATus:CALHist:10 Entry 10 (oldest)

Answer:Index, cal-factor, phase, leak rate, year/month/
day, hour:min:sec, INT/EXT/MAN, gas type, p1, p2, on
time, probeID, S-value, temperature, cal state
Example: 05 Fac:1.32E+01 Phase:35 Leak:3.59E+00
2013/02/07 15:23:25 EXT Smart R134a p1:+324 p2:-396
t_on:002 ProbeID:D535 S-value:550 T:+27.2 CalState:23

275

R

*STATus:ERRHist

Error history
*STATus:ERRHist Actual entry
*STATus:ERRHist:1 Entry 1 (newest)
*STATus:ERRHist:2 Entry 2
...
*STATus:ERRHist:15 Entry 15 (oldest)

287

R

*STATus:ERRor

current number of error / warning („NO ERROR/
WARNING“ or "ERROR " and 3-digit failure number)

290

R

*STATus:ERRor:VALue

actual value according to the actual error / warning

289

R

Command

Meaning

Relates
to LD
cmd. no.

Read /
Write