INFICON HLD6000 Refrigerant Leak Detector Interface Description User Manual

Page 24

Advertising
background image

24

LD Protocol

297

129

Status Present warnings

R

UINT32

Each bit represents a warning
Bit 0 = Warning main EEPROM
Bit 1 = Warning IR source
Bit 2 = not used
Bit 3 = not used
Bit 4 = WARNING_MAIN_TEMP
Bit 5 = WARNING_CAL_REQUEST
Bit 6 = WARNING_RTC
Bit 7 = WARNING_PROBE_COM
Bit 8 = WARNING_DIFF_PRESS
Bit 9 = WARNING_IO_DISCONNECT
Bit 10 = WARNING_PROBE_SUPPLY
Bit 11 = WARNING_I_MOD
Bit 12 = WARNING_PROBE_NO_ANSWER
Bit 13 = WARNING_PROBE_PARA
Bit 14 = WARNING_LEAK_EMPTY
Bit 15 = WARNING_LEAK_NEARLY_EMPTY
Bit 16 = WARNING_LEAK_EEPROM
Bit 17 = WARNING_LEAK_TEMP
Bit 18 = WARNING_LIGHT_BARRIER
Bit 19 = WARNING_AUDIO_AMPLIFIER
Bit 20 = WARNING_12V_OUT_OF_RANGE
Bit 21 = WARNING_12VN_OUT_OF_RANGE
Bit 22 = WARNING_5V_OUT_OF_RANGE
Bit 23 = WARNING_CU_DISCONNECT
Bit 24 = WARNING_3V3_OUT_OF_RANGE
Bit 25 = WARNING_24V_OUT_OF_RANGE
Bit 26 = WARNING_NO_ANSWER_BUS_MOD
Bit 27 = WARNING_LAMP_VOLTAGE

300

12C

Status Device identification

R

UINT8[2]

Device identification, always {1,50} for HLD6000

301

12D

Status Device name

R

CHAR[*]

Get device name as ASCII string, always "HLD6000"

302

12E

Status Sniffer probe type

R

UINT8

0: probe not detected
1: HLD5000-probe
2: HLD6000-probe

310

136

Status SW-version main board

R

UINT8[3]

Software version
Index 0: Main version
Index 1: Sub version
Index 2: Debug version

311

137

Status SW-version probe

programming master

R

UINT8[3]

Software version of master, by which this sniffer probe was
programmed
Index 0: Main version
Index 1: Sub version
Index 2: always 0

312

138

Status SW-version sniffer probe

R

UINT8[3]

Software version probe
Index 0: Main version
Index 1: Sub version
Index 2: always 0

Command