2 normal protocol, Normal protocol – INFICON HLD6000 Refrigerant Leak Detector Interface Description User Manual

Page 5

Advertising
background image

Interface Protocols

5

Now: Last 3 digits are error number, no message type

1.2.2

Normal Protocol

Command ‘T’

Command ‘T’ not supported

Reason: Totally different display / menu at the HLD6000

Display IDs

03: Program menu

Reason: Totally different menu access control at the HLD6000

Parameter IDs

2: Function of the probe button

‘3 = SEARCH’ omitted; no search-mode in HLD6000, use the second set

point

20: Baud Rate

Reason: Fixed baud rate in HLD6000

35: PIN-Code

Reason: Totally different menu access control at the HLD6000

55: Contrast

Reason: No Contrast setting for HLD6000 display

56: Language

Reason: Different language setting, no influence on device behavior via

serial interface

Remote IDs

00: Enable program menu

Reason: Totally different menu access control at the HLD6000

01: Lock program menu

Reason: Totally different menu access control at the HLD6000

11…18: Simulates a pressed Softkey

Reason: No Softkeys at HLD6000 front panel

Status IDs

67: Last error in error list

Changed: ‘T’ and ‘NN’ are used for the 3-digit error number

79: Current status of the LED display

Reason: No LED display in HLD6000

81: Menu status

Reason: Totally different menu at the HLD6000

Advertising