3 status word, Status word, Status – INFICON HLD6000 Refrigerant Leak Detector Interface Description User Manual
Page 19
Advertising
LD Protocol
19
3.3
Status Word
Status word bit no.
Meaning
Bit 0
Device state:
0 = Runup
1 = Standby
2 = Measure
3 = Calibration internal
4 = Calibration external
5 = Proof
6 = Not used
7 = Not ready
Bit 1
Bit 2
Bit 3
not used
Bit 4
not used
Bit 5
Still pending device warning. Is 1 even if
warning is confirmed via command "Clear
error" (Command No. 5) as long as
reason for warning still exists.
Bit 6
Sniffer Key
Bit 7
not used
Bit 8
PLC Output Change
Bit 9
Setpoint exceeded
Bit 10
Actual valid setpoint:0=Setpoint 1,
1=Setpoint 2
Bit 11
Value changed via interface command
Bit 12
Light barrier
Bit 13
Unconfirmed device warning
Bit 14
Device error
Bit 15
Syntax/Command error