LD Protocol

19

3.3

Status Word

Status word bit no.

Meaning

Bit 0

Device state:
0 = Runup
1 = Standby
2 = Measure
3 = Calibration internal
4 = Calibration external
5 = Proof
6 = Not used
7 = Not ready

Bit 1
Bit 2

Bit 3

not used

Bit 4

not used

Bit 5

Still pending device warning. Is 1 even if
warning is confirmed via command "Clear
error" (Command No. 5) as long as
reason for warning still exists.

Bit 6

Sniffer Key

Bit 7

not used

Bit 8

PLC Output Change

Bit 9

Setpoint exceeded

Bit 10

Actual valid setpoint:0=Setpoint 1,
1=Setpoint 2

Bit 11

Value changed via interface command

Bit 12

Light barrier

Bit 13

Unconfirmed device warning

Bit 14

Device error

Bit 15

Syntax/Command error