INFICON HLD6000 Refrigerant Leak Detector Interface Description User Manual

Page 27

Advertising
background image

LD Protocol

27

438

1B6

Param PLC input configuration IO

module

R/W UINT8[10] Configuration of PLC input port of the IO module

Index 0...9 = PLC_IN1...PLC_IN10
See table
-6 = VALID_TRIGGER_N
-5 = START_STOP_N
-4 = CLEAR_N
-3 = STOP_N
-2 = START_N
-1 = CAL_EXTERN_N
0 = NO_FUNCTION
1 = CAL_EXTERN
2 = START
3 = STOP
4 = CLEAR
5 = START_STOP
6 = VALID_TRIGGER
Default:
PLC_IN 1: NO_FUNCTION
PLC_IN 2: NO_FUNCTION
PLC_IN 3: START_STOPPLC_IN 4: NO_FUNCTION
PLC_IN 5: CAL_EXTERN
PLC_IN 6: NO_FUNCTION
PLC_IN 7: CLEAR
PLC_IN 8: NO_FUNCTION
PLC_IN 9: NO_FUNCTION
PLC_IN 10: NO_FUNCTION

440

1B8

Meas

Remaining life time COOL-
Check [d]

UINT16

Days until internal reference leak is empty

450

1C2

Param Date+Time [YMDhms]

R/W UINT8[6]

Date and time
use only with array-index 255 (all bytes)
year (1..99), month, day,
hour (0..23), min, sec

459

1CB Param Phase

R/W UINT8

Phase (Min: 0; Max:39)

460

1CC Param Calibration factor instrument

R/W FLOAT

Calibration factor (Min: 6, Default: 15, Max.: 80)

480

1E0

Param Auto standby interval [min]

R/W UINT8

Auto standby interval 0 = OFF

485

1E5

Param User gas factors

R/W FLOAT[3]

Gas factors for user gases:
Index 0: User gas 1
Index 1: User gas 2
Index 2: User gas 3

486

1E6

Param Gas type (smart probe only)

R/W UINT8

Gas type (smart probe only)
5 = R22
22 = R404A
23 = R407C
24 = R410A
25 = R134A
26 = HFO1234YR
27 = R32
40 = USER-GAS 1
41 = USER-GAS 2
42 = USER-GAS 3

489

1E9

Param ID of last calibrated probe

R

UINT16

ID of last calibrated probe (Min.: 0; Default: 2; Max.: 15)

490

1EA

Param Probe ID

R

UINT16

ID of actual probe

Command

Class Name

R/W Data type Description