INFICON HLD6000 Refrigerant Leak Detector Interface Description User Manual

Page 26

Advertising
background image

26

LD Protocol

413

19D

Param Sniffer LED alarm configuration R/W UINT8

Configures the behavior of the white sniffer pobe LEDs in
case of trigger alarm:
0=disabled (same Brightness as during normal
measurement)
1=on (higher brightness as during normal measurement if
possible)
2=blink
(Default: 2)

414

19E

Param Sniffer white LED brightness

R/W UINT8

Configures the brightness of the white sniffer pobe LEDs in
case of normal measurement:
0=off ... 6=max Brightness
(Default: 4)

418

1A2

Param Cal. request interval

R/W UINT16

Calibration request interval [min.]
(Min: 0; Default: 60; Max: 1440)

420

1A4

Param Volume

R/W UINT8

Volume of Triggeralarm
(Min.: 0; Default: 5; Max.: 15)

422

1A6

Param Probe key config

R/W UINT8

Configuration of probe key
0: probe button not used for trigger switching
1: probe button switches between trigger1 and trigger 2
(Default: 1)

423

1A7

Param Speaker beep

W

UINT8[2]

Performs a speaker beep.
Index 0: Typ
Index 1: Volume (0..15)
Typ: 1=2000Hz, 2=first 2000Hz, then 1000Hz

424

1A8

Status Time until next calibration

request [s]

R/W UINT32

Time until next calibration request in seconds

425

1A9

Param Activate Calibration request

R/W UINT8

0 = deactivate a calibration request
1 = activate a calibration request
(Default: 0)

432

1B0

Param Leak rate unit

R/W UINT8

Leak rate unit
0 - g/a
1 - lb/yr
2 - mbarl/s
3 - oz/yr
4 - Pam3/s
(Default: g/a)

Command

Class Name

R/W Data type Description

dez

hex