3 commands, Commands – INFICON HLD6000 Refrigerant Leak Detector Interface Description User Manual

Page 8

Advertising
background image

8

ASCII Protocol

2.3

Commands

Command

Meaning

Relates
to LD
cmd. no.

Read /
Write

*BEEP

Output beep sound. Parameter: Typ,Volume
"*BEEP 1,2"
Typ: 1= 2000 Hz, 2=Dual (2000 and 1000 Hz), 3= 1000 Hz
Volume: Volume between 1 and 15

W

*CAL

Start external calibration

4

W

*CAL:ACKnowledge

Acknowledge stable signal during external calibration

11

W

*CLS

Clear Error

5

W

*CONFig:AUDio

Audio alarm type (PIN, SET, TRIG, OFF)

600

R/W

*CONFig:BRIGHTness

Sniffer white LED brightness

414

R/W

*CONFig:BUTSniffer

button of the sniffer probe (OFF, ON)

422

R/W

*CONFig:CALleak

leak rate of external test leak

R/W

*CONFig:CALleak:G/a

leak rate of external test leak in g/a

388

R/W

*CONFig:CALleak:LB/yr

leak rate of external test leak in lb/yr

R/W

*CONFig:CALleak:MBAR*l/s

leak rate of external test leak in mbar*l/s

R/W

*CONFig:CALleak:OZ/yr

leak rate of external test leak in oz/yr

R/W

*CONFig:CALleak:PA*m3/s

leak rate of external test leak in Pa*m3/s

R/W

*CONFig:CALREQInterval

calibration request interval in minutes (0=calibration
request off)

418

R/W

*CONFig:GAS

Gas type (smart probe only)
(R134A, R22, R404A, R407C, R410A, R1234YF, R32,
USER1, USER2, USER3)

486

R/W

*CONFig:LIGHTAlarm

Sniffer white LED alarm configuration:
DISABLED: always same brightness
ON: higher brightness if trigger 1 is exceeded
BLINK: blink, if trigger 1 is exceeded

413

R/W

*CONFig:PLCINLINK

Assignment of PLC-inputs
*CONFig:PLCINLINK:1 for PLC input Pin1
*CONFig:PLCINLINK:2 for PLC input Pin2
...
*CONFig:PLCINLINK:10 for PLC input Pin10

438

R/W

*CONFig:PLCINLINK:1

Configuration of PLC-input. The following settings are
possible: NOT_USED, EXT_CAL, START, STOP,
CLEAR, START_STOP, VALID_TRIGGER.

438

R/W

*CONFig:PLCINLINK:n

n = 2…10 see CONFig:PLCINLINK:1

438

R/W

*CONFig:PLCOUTLINK

Assignment of PLC-outputs
*CONFig:PLCOUTLINK:1 for Pin 1
*CONFig:PLCOUTLINK:2 for Pin 2
...
*CONFig:PLCOUTLINK:8 for Pin 8

263

R/W

*CONFig:RECorder:LINK1

Function at analog output channel 1
(OFF, LR_LIN, EXTERN)

222

R/W

*CONFig:RECorder:LINK2

Function at analog output channel 2
(OFF, LR_LIN, EXTERN)

222

R/W

*CONFig:RECorder:UPPERVAL

Upper limit for the analog out at IO module.
The value sets the leak rate in g/a for an output voltage of
10V

225

R/W

*CONFig:RS232

Protocol (ASCII, LD, HLD4000, SIMPLE)

26

R/W

Advertising