4 examples, 5 error messages, Examples – INFICON HLD6000 Refrigerant Leak Detector Interface Description User Manual
Page 15: Error messages

ASCII Protocol
15
2.4
Examples
External Calibration
To execute an external calibration via ASCII-protocol proceed as follows:
1 Move sniffer to external leak
2 Start calibration with command *CAL:EXT
3 Wait until leack rate signal is stable
4 Acknowledge stable signal with command *CAL:ACK
5 Poll *STATUS:CAL? Until it answers "RESULT_OK"
6 Remove sniffer from external leak
2.5
Error Messages
Command
answer
*stat? (CR)
MEAS (CR)
mode
*status? (CR)
MEAS (CR)
mode
*read? (CR)
2.876E-7 (CR)
leak rate according to
programmed unit
*read:pa*m3/s? (CR)
2.876E-6 (CR)
leak rate in a different unit
*start (CR)
OK (CR)
Exit stand by mode; start
measurement
*conf:trig1? (CR)
1.0E-9 (CR)
retrieve trigger 1
*conf:trig1 2.0E-9 (CR)
OK (CR)
set trigger 1
Message
Meaning
OK
command completed
E01
wrong command start (no „*“)
E02
illegal blank
E03
command word 1 illegal
E04
command word 2 illegal
E05
command word 3 illegal
E06
control by RS232 not enabled
E07
argument faulty
E08
no data available
E09
error buffer overflow
E10
command actually invalid
E11
query not allowed
E12
only query allowed
E13
command not implemented