ASCII Protocol

15

2.4

Examples

External Calibration

To execute an external calibration via ASCII-protocol proceed as follows:

1 Move sniffer to external leak

2 Start calibration with command *CAL:EXT

3 Wait until leack rate signal is stable

4 Acknowledge stable signal with command *CAL:ACK

5 Poll *STATUS:CAL? Until it answers "RESULT_OK"

6 Remove sniffer from external leak

2.5

Error Messages

Command

answer

*stat? (CR)

MEAS (CR)

mode

*status? (CR)

MEAS (CR)

mode

*read? (CR)

2.876E-7 (CR)

leak rate according to
programmed unit

*read:pa*m3/s? (CR)

2.876E-6 (CR)

leak rate in a different unit

*start (CR)

OK (CR)

Exit stand by mode; start
measurement

*conf:trig1? (CR)

1.0E-9 (CR)

retrieve trigger 1

*conf:trig1 2.0E-9 (CR)

OK (CR)

set trigger 1

Message

Meaning

OK

command completed

E01

wrong command start (no „*“)

E02

illegal blank

E03

command word 1 illegal

E04

command word 2 illegal

E05

command word 3 illegal

E06

control by RS232 not enabled

E07

argument faulty

E08

no data available

E09

error buffer overflow

E10

command actually invalid

E11

query not allowed

E12

only query allowed

E13

command not implemented