LD Protocol

Command info

Data types

All data types are used in Big Endian format (Motorola format), i.e. the byte with the
highest-order bits is transferred first.

Arrays

• Read single elements: Array index in first DATA-byte

• Write single elements: Array index in first DATA byte and values in following

DATA bytes

• Read all elements: Pseudo array index 255 in first DATA byte

• Write all elements: Pseudo array index 255 in first DATA byte and values in

following DATA bytes

• Response from slave (in case data are sent): Array index or pseudo array index

in first DATA byte and values in following DATA bytes

All elements of an array have the same Min/Def/Max value.

Array parameters in commands table (see chapter

3.4

): The number of array

elements is set in brackets behind the data type.

1. Byte

Data type (see table “Data types”)

2. Byte

Number of array elements:
0 = no data, no array
1 = data, no array
2 ... 255 = array

3. Byte

Bit 0: 1 = Reading allowed, 0 = Reading not allowed
Bit 1: 1 = Writing allowed, 0 = Writing not allowed
Bit 2 ... 7: always 0 (not used)

Value Meaning

Acronym

Comments

1

Signed 8 bit integer

SINT8

2

Signed 16 bit integer

SINT16

3

Signed 32 bit integer

SINT32

4

Unsigned 8 bit integer

UINT8

5

Unsigned 16 bit integer

UINT16

6

Unsigned 32 bit integer

UINT32

7

Character

CHAR

ISO 8859-1; printable characters

16

Signed 64 bit integer

SINT64

17

Unsigned 64 bit integer

UINT64

18

Floating point/real number FLOAT

4 Byte single float according to
IEEE 754

20

no data

NO_DATA For commands without data, such