INFICON HLD6000 Refrigerant Leak Detector Interface Description User Manual

Page 28

Advertising
background image

28

LD Protocol

600

258

Param Audio alarm type

R/W UINT8

Audio alarm type
0 = Alarm off
1 = Trigger alarm
2 = Setpoint
3 = Pinpoint
(Default: Setpoint)

1161 489

Control Parameter reset

W

UINT8

Parameter reset:
0: Load factory settings

1581 62D

Meas

Acceleration

R

SINT16[3] [0] = X-value

[1] = Max (Y & Z)-value
[2] = 0

1631 65F

Meas

Light barrier voltages [V]

R

FLOAT[2]

Light barrier voltages at photodiode
Index 0: LED is off
Index 1: LED is on

1800 708

Status Active protocol IO

R

UINT8

Active interface protocol for I/O module. Defined by DIP
switch at I/O module or command 2593.
Values: See command 2593

1815 717

Status Reset source

R

UINT8

Shows the last reason of reset

2593 A21

Param Interface protocol IO

R/W UINT8

Selected interface protocol for I/O module. May be
overwrite if DIP switch at I/O module are not set to "000"
0 = LD
1= ASCII
don't use following settings for new constructions
8= 'Normal' of the HLD5000
9 = 'Simple' of the HLD500
(Default: ASCII)

2659 A63

Param Sniffer tip filter change request R/W UINT8

Sniffer tip filter change request:
0=0ff, 1=on
(Default: on)

Command

Class Name

R/W Data type Description

dez

hex

Advertising