INFICON HLD6000 Refrigerant Leak Detector Interface Description User Manual
Page 28

28
LD Protocol
600
258
Param Audio alarm type
R/W UINT8
Audio alarm type
0 = Alarm off
1 = Trigger alarm
2 = Setpoint
3 = Pinpoint
(Default: Setpoint)
1161 489
Control Parameter reset
W
UINT8
Parameter reset:
0: Load factory settings
1581 62D
Meas
Acceleration
R
SINT16[3] [0] = X-value
[1] = Max (Y & Z)-value
[2] = 0
1631 65F
Meas
Light barrier voltages [V]
R
FLOAT[2]
Light barrier voltages at photodiode
Index 0: LED is off
Index 1: LED is on
1800 708
Status Active protocol IO
R
UINT8
Active interface protocol for I/O module. Defined by DIP
switch at I/O module or command 2593.
Values: See command 2593
1815 717
Status Reset source
R
UINT8
Shows the last reason of reset
2593 A21
Param Interface protocol IO
R/W UINT8
Selected interface protocol for I/O module. May be
overwrite if DIP switch at I/O module are not set to "000"
0 = LD
1= ASCII
don't use following settings for new constructions
8= 'Normal' of the HLD5000
9 = 'Simple' of the HLD500
(Default: ASCII)
2659 A63
Param Sniffer tip filter change request R/W UINT8
Sniffer tip filter change request:
0=0ff, 1=on
(Default: on)
Command
Class Name
R/W Data type Description
dez