Trouble Shooting

5.3

LD Protocol specific

Error

Possible Reason

Solution

IO1000 does not reply

Wrong Address

Always use Address 1 in LD protocol.

Other protocol errors

Try to use NOP command (05hex 04hex 01hex
00hex 00hex 77hex) first, to check if connection
works in general. The answer should be 02hex
05hex XXhex XXhex 00hex 00hex XXhex

IO1000 replys with CRC
error (error code 1)

Wrong CRC calculation

Check you CRC code calculation. See example C
source file "CRC_calculation.c" provided by
INFICON. Check your code with unit test function in
this source code file.