ASCII Protocol

*STATus:WARNINGBits

Warning-Bits which indicate pending warnings:
WARNING_MAIN_EEPROM = 0x00000001
WARNING_LAMP = 0x00000002
WARNING_MAIN_TEMP = 0x00000010
WARNING_CAL_REQUEST = 0x00000020
WARNING_RTC = 0x00000040
WARNING_PROBE_COM = 0x00000080
WARNING_DIFF_PRESS = 0x00000100
WARNING_IO_DISCONNECT = 0x00000200
WARNING_PROBE_SUPPLY = 0x00000400
WARNING_I_MOD = 0x00000800
WARNING_PROBE_NO_ANSWER = 0x00001000
WARNING_PROBE_PARA = 0x00002000
WARNING_LEAK_EMPTY = 0x00004000
WARNING_LEAK_NEARLY_EMPTY = 0x00008000
WARNING_LEAK_EEPROM = 0x00010000
WARNING_LEAK_TEMP = 0x00020000
WARNING_LIGHT_BARRIER = 0x00040000
WARNING_AUDIO_AMPLIFIER = 0x00080000
WARNING_12V_OUT_OF_RANGE = 0x00100000
WARNING_12VN_OUT_OF_RANGE = 0x00200000
WARNING_5V_OUT_OF_RANGE = 0x00400000
WARNING_CU_DISCONNECT = 0x00800000
WARNING_3V3_OUT_OF_RANGE = 0x01000000
WARNING_24V_OUT_OF_RANGE = 0x02000000
WARNING_NO_ANSWER_BUS_MOD = 0x04000000

297

R

*STOp

Switch from Measure to Standby

2

W

Command

Meaning

Relates
to LD
cmd. no.