4.3.2

Write Process Data (Leak Detector

PLC)

These 29 data bytes are send periodically from the leak detector to the field bus master (e.g. a programmable logic controller):

Attention: PROFIBUS and PROFINET IO sends high byte first, DeviceNet and EtherNet/IP sends low byte first.

Title

Byte

Bit

Name

Meaning

Similar to IO1000

Output

Similar to RS232

ASCII cmd.

Similar to RS232

LD cmd.

status word

1

(low

byte)

0

Device state

0=Run up

1=Standby

2=Measure

3=Calibration internal

4=Calibration external

5=Proof

6=Not used

7=Not ready

Run up, CAL active,

Proof active, Error,

Measure

*STATUS?

status word

1
2

3

not used

4

not used

5

Still pending device

warning

Warning

*STATUS:WARNINGBITS?

6

Sniffer key

0 = Sniffer probe key not pressed

1 = Sniffer probe key pressed

Sniffer button

*STATUS:SNKEY?

7

not used

2

(high
byte)

0

Calibration request

active

1= Leak detector requests a calibration

CAL_REQUEST

*STATUS:WARNINGBITS?

1

Setpoint

Trigger1, Trigger2

*STATUS:TRIGGER?

2

Active setpoint

3

not used

4

Light barrier

Light barrier

5

Unconfirmed device

warning

6

Device error

Error

*STATUS:ERROR?

7

not used