LD Protocol
29
3.5
Error Messages
Telegram error handling
• Slave discards all characters until it receives a ENQ as telegram start identifier.
• Slave does not generate an error message, if address is not correct.
• Slave reports CRC errors with error message 1 (CRC failure)
• Slave reports length errors with error message 2 (Illegal telegram length) or 11
(Data length is not correct for the command)
To prevent the response from colliding with the next request, the slaves do not
respond in case of a timeout.
Error numbers (if status word Bit 15 is set 1)
In case of error: STX, LEN, Stw, Cmd and one Data-Byte (with error number) sent
Error No. Meaning
1
CRC-failure
2
Illegal telegram length
10
command doesn't exist
11
Data length is not correct for the command
12
Read not allowed
13
Write not allowed
14
Array-Index out of range or missing
20
Control actually not allowed with this interface
21
Password not OK
22
Command actually not allowed (e.g. calibration during Run-Up)
30
Data not in range
31
No data available