EN- 101
Sensistor ISH2000 Technical Reference Manual (TRM)
4. When changing the PCB the
cables must be removed by
soldering.
Figure E-30:
5. To replace the probe tube release
the safety screw by using an Allen
key. Heat the probe connection
with a warm gun lift and loosen the
probe tube. At assembly the probe
tube must be bonded with Loctite.
Figure E-31:
Note: When mounting, ensure that
the nut on the Lemo connector on
the the PCB is properly adjusted
(distance shall be 9 mm from
LEMO's edge to the nut).
E 411