Sensistor ISH2000 Technical Reference Manual (TRM)

EN- 42

Probe type determines information printed.

Analysis data output

The Analysis Data Output option is intended for transferring test results to a
supervising computer system such as, for example, a PLC system.

Communication specifications.

The data format for Analysis data output consists of nine ASCII characters.
Seven characters show the value in engineering format one character shows
the result of the test, and one character shows line feed (LF).

The character indicating the result of the test is one of the following.

Probe type

Data printed

Hand Probe P50

1, 2, 3, 4, 6

Counter Flow Hand Probe
AP57

1, 2, 3, 4, 6

Sniffer Hand Probe AP55

1, 2, 4, 5, 6

Sampling Units AP29 and
H28

1, 2, 4, 5, 6

APC driver with active test
(using MEAS flag)

1, 2, 4, 5, 6

APC driver not using
MEAS flag

1, 2, 3, 4, 6

Setting

Value

Data rate

9600 baud

Data bits

8

Stop bits

1

Parity

None

Flow control

None

Character

Result of the test

A

Accept. Previous test was below Reject Level alarm limit.

R

Reject. Previous test was above Reject Level alarm limit.

P

Rejected by Purging. Previous test was above purge limit
(and Reject level limit).

C

Calibration approved. Previous cycle was calibration.
Calibration was approved.

F

Calibration failed. Previous cycle was calibration.

E

Test interrupted by “Error” that occurred during cycle
(probe or sensor error etc.).