Sensitivity fault analysis, E 130, Figure e-4: sensitivity fault analysis – INFICON Sensistor ISH2000 Hydrogen Leak Detector User Manual

Page 83

Advertising
background image

EN- 83

Sensistor ISH2000 Technical Reference Manual (TRM)

E 130

h. Acceptance level for new sensor is minimum 1.2 = 1.20E+00.

i. If the Signal is below 1.2: Wait at least 10 minutes (leave instrument on)
and then repeat the test. If the Signal (after a repeated test) is still below
1.2 the sensor is to be considered no good for delivery.

Sensitivity fault analysis

The following questions below are useful to detect a problem.

Figure E-4: Sensitivity fault analysis

Type of

problem?

Quick or

slow

signals?

Electrical fault.
Bad cable or
connector, or
earthing problem .
This is most
likely a handling
error by
customer.

Check for gas spillage
in the background.
(could be exhausts,
smoke, tracer gas or
gastric gases ) .

This is an application
problem. Not a
warranty issue.

Low sensitivity

False signals

>5 sec .

< 1 sec.

Detector

message:

“Check

Sensor,

Voltage Error”

Yes

No

1 -
5

sec.

No explanation

No explanation

* Service Mode Signal is read in Service Mode. See instructions for Sensitivity Test

Frequent unexplained
Service Mode Signals*
above 0.18 that can be
confirmed at the service
center are grounds for
warranty replacement.

The Reject level
should have been
designed to 10 ppm
H2 or more.
For Reject levels
it is normal to change
sensor 1-2 times per
year.

Detector

message:

“ Check

Probe and

Cable?”

Yes

Detector

message:

“ Check

Sensor, Temp”

Yes

Check that the sensor
is properly connected
to the probe. Refer to
D 200 General trouble
-shooting guide.

Check that the sensor
is properly connected
to the probe.Refer to
D 200 General trouble
-shooting guide

Check that the probe
cable is properly
connected to the probe
and the instrument.