D 200 general troubleshooting guide, Symptoms description, D 200 – INFICON Sensistor ISH2000 Hydrogen Leak Detector User Manual

Page 74

Advertising
background image

Sensistor ISH2000 Technical Reference Manual (TRM)

EN- 74

D 200

D 200 General troubleshooting guide

Symptoms description

Fault symptoms

Fault

Measures

Error messages: “Check
Probe and Cable
”. Red
LED flashes quickly.

Broken connection
between probe and
instrument

Check that the probe
cable is properly con-
nected to the probe and
the instrument. If the fault
persists, replace the
probe/cable.

Software problems

Install the software

Broken Main PCB

Replace Main PCB

Broken Probe PCB

Replace Probe PCB

Error messages: “Error

Error in Active Probe

See Probe Manual.

Error messages: “Check
Sensor, Voltage Error

Sensor voltage from the
gas sensor out of the
range 300 - 3500 mV

Check that the sensor is
properly connected to the
probe. If the fault persists,
replace the sensor.

Error messages: “Check
Sensor, Temp.

Diode voltage from the
gas sensor out of the
range 544 - 612 mV

Check that the sensor is
properly connected to the
probe. If the fault persists,
replace the sensor.

No sound in Detection
Mode and
Analysis Mode.

Sound set low

Press the + button repeat-
edly.

Broken speaker

Replace the speaker

No picture on display, no
sound

Fuse broken

Check the fuse.

Broken Main Cable

Replace the Main Cable

Discharged battery

Broken Main PCB

Replace the Main PCB

No picture but only sound
when exposed to gas.

Display setting may be
wrong

Watch the display from
the side at a low angle
and aim a lamp at the
screen. Try to see the text
so that you can enter the
Display Settings menu and
adjust contrast and
brightness.

Display or Inverter Cable
loose

Reconnect loose wiring

Display Lamp broken

Replace the display

Loss of spots or black lines
on screen

Internal contact to LCD
broken

Replace the display

Bad sound out of the
speaker

Broken speaker

Replace the speaker