Sensistor ISH2000 Technical Reference Manual (TRM)

B 310 Controlling the detector with a PC

computer through the RS 232 interface

RS 232 interface

The RS 232 interface makes it possible to control the leak detector with a PC
compatible computer: it is installed on all leak detectors.

Commands available for your leak detector

RS232 interface commands

Common used functions

K = Calibration request

Starts calibration if the Sensistor ISH2000 has an active probe driver installed.
Sensistor ISH2000 answers with a "K" if an active driver containing a calibra-
tion routine was found and "F" if the calibration APC sequence not found. Cal-
ibration doesn’t start if Purge level is reached.

M = Measure Request

The active test cycle defined by the APC driver starts. “M” is returned if the
selected driver supports active test. “F” (failed) is returned otherwise.

N = Print Request

Returns current analysis value.

Setting

No printer

PC printer

Data output

Data rate

115200 baud

1200 baud

9600 baud

Data bits

8

8

8

Stop bits

1

1

1

Parity

None

None

None

Flow control

None

None

None

Command

Header

Calibrate

K

Measure

M

Print Request

N

Stop Measurement

Q

Hand Probe

R

Active Probe (Installed AP)

S

Analysis Mode

X

Detection Mode

Z

Combined Mode

Y