Gas sensing specification (in analysis mode), A 900 – INFICON Sensistor ISH2000 Hydrogen Leak Detector User Manual

Page 25

Advertising
background image

EN- 25

Sensistor ISH2000 Technical Reference Manual (TRM)

A 900

Gas sensing specification (In Analysis Mode)

Selected unit

Sensitivity

Measurement
range

Linearity

Repeatability

ppm (H

2

)

0.5 ppm

0.5 - 2000 ppm
(0.2%)

Typ. ± 15% of reading
(within 0.1 - 10 x
calibration point in
range 0.5 - 100 ppm)

Typ. ± (10% of
reading + 0.3 ppm)

mbarl/s air
(using 5% H

2

/

95% N

2

as

Tracer Gas)

5 x 10

-7

mbarl/s

5 x 10

-7

- 4 x 10

-2

mbarl/s

Typ. ± 15% of reading
(within 0.1 - 10 x
calibration point in

range 1 x 10

-5

- 2 x

10

-

3

mbarl/s)

Typ. ± (10% of

reading + 3 x 10

-7

mbarl/s)

g/a R143a (using
5% H

2

/95% N

2

as Tracer Gas)

0.2 g/a

0.2 - 8300 g/a

Typ. ± 15% of reading
(within 0.1 - 10 x
calibration point in
range 0.2 - 420 g/a

Typ. ± (10% of
reading + 0.1 g/a)