Status signal patterns, B 300 – INFICON Sensistor ISH2000 Hydrogen Leak Detector User Manual

Page 35

Advertising
background image

EN- 35

Sensistor ISH2000 Technical Reference Manual (TRM)

B 300

Status signal patterns

Status signals for pin 14 - 18.

Figure B-5: Example screen

Where:

DET_ERROR will go high for a short time (1-5 seconds) when the detector is
switched on. It will go low when the sensor has been checked. In normal oper-
ation, DET_ERROR = HIGH means that there is a problem with the sensor,
probe, or cable.

DET_WAIT is high when instrument is in warm-up mode after switching on
power. Instrument will also go into warm-up if there is a temporary fault in the
sensor or sensor connection.

The timing of the status signals in relation to different events is described by the

following two examples:

• Example: Input signals issued to control the APC system should have a pulse
length of at least 40 ms.

• Example: Output signals switch with a cycle time of 20 ms (0.02 s). This is the
cycle time of the APC system.

Note: Not valid for battery operated version of Sensistor ISH2000.

I/0 name on the screen

Signal Name on the pins

Function

Inputs 0 to 4

In_0 to In_4

Input signal

Outputs 0 to 6

Out_0 to Out_6

Output signal

Status O

DET_ON

When detector is on

Status W

DET_WAIT

High during warm-up

Status S

DET_SIGNAL

Gas detected / Sensor not
recovered

Status R

LEAK_ALARM

Leak above Reject Level
detected

Status C

CAL_CONF

Calibration preformed
and ok

Status E

DET_ERROR

High if Probe, Sensor or
Cable is broken