Service settings, C 201 – INFICON Sensistor ISH2000 Hydrogen Leak Detector User Manual

Page 53

Advertising
background image

EN- 53

Sensistor ISH2000 Technical Reference Manual (TRM)

C 201

means that no password is needed to modify the critical parameters. The
default is no password (“”).

Contact INFICON AB if you have lost your user password. If the Password Pro-
tected Calibration parameter is set to ON you will be prompted for a password
when starting a calibration.

Note: Setting Password Protected Calibration to ON has no effect if no pass-
word is set.

Note: APC controlled calibration can be started from the bus in both cases.

Audio Base Frequency

This sets the lowest audio base frequency tone in Search and Detection Mode.

Set Clock

Real time set as hh:mm:ss. Hours and minutes can be adjusted. Seconds will
automatically be set to 00 when hours and minutes have been set. Clock runs
even when detector is disconnected from the power supply.

Set Date

Real Time Clock date set as YY-MM-DD. Clock runs even when detector is dis-
connected from the power supply.

Printer Port

The Sensistor ISH2000 is equipped with a serial (RS232) printer port. See
“Printer port” on page 35.

Info

Contains information about software versions, Serial number, and Internet con-
tact information.

Service settings

The Service Mode is reached by starting the instrument and at the same time
hold the right button down on the panel. After start a new main menu called
Service Settings will appear.

See “E 140 Service menu” on page 84