C 304 correlation value, C 304 – INFICON Sensistor ISH2000 Hydrogen Leak Detector User Manual

Page 62

Advertising
background image

Sensistor ISH2000 Technical Reference Manual (TRM)

EN- 62

C 304

C 304 Correlation value

If you want to use another unit you have to put a recalculation number into
Correlation Value which describes the relationship between the different
units.

This might be necessary when you want to display Reject Level unit other than
the reference leak rate unit.

Useful also for correct the value when volume and time affect the measure
value result in accumulation test.

Example:

To use mbarl/s as an calibrate reference unit and mm3/s as a analysis Unit you
have to set a Correlation Valus to 1.00E-03.