PN

07

4-

61

3-

P1

A

STM-2 Operating Manual

2.4 STM-2 Indicators

2.4.1 Power

Illuminated . . . . . . . . . . . . . . . . . . . . STM-2 is powered up and is connected to a

good crystal.

The host computer has initialized STM-2 by
returning a reset status when the software
begins communication.

Flashing fast . . . . . . . . . . . . . . . . . . . STM-2 is powered up but cannot detect a

good crystal.

Blinking slow
(approximately once per second) . . . STM-2 is powered up and the crystal is good,

but the computer application has not
initialized STM-2. Once the computer
initializes STM-2 the indicator will illuminate
continuously (see

section 6.1.1 on page 6-1

).

Extinguished. . . . . . . . . . . . . . . . . . . STM-2 does not have power. Check the USB

connection, and make sure the computer is
turned on.

2.4.2 USB

The USB indicator detects communications signal traffic.

Illuminated . . . . . . . . . . . . . . . . . . . . STM-2 is connected and communicating to a

host computer. Communications
sent/received every 100 ms will cause the
indicator to be steadily illuminated.

Flashing . . . . . . . . . . . . . . . . . . . . . . STM-2 is connected and communicating to a

host computer. A flashing indicator will
correspond to the time elapsed between sent
and received communications. A query sent
once per second will correspond to the
indicator flashing approximately once per
second.

Extinguished. . . . . . . . . . . . . . . . . . . STM-2 is not communicating to the

computer.

NOTE: STM-x_win32 LabVIEW application software and INFICON STM-2

software will steadily illuminate the indicator due to communications
queries sent every 100 ms. User-created software may not steadily
illuminate the indicator due a longer time period elapsing between
communications queries being sent.