STM-2 Operating Manual

Zeroing commands

B

None

B

Zeroes timer and thickness

C

None

C

Zeroes thickness

D

None

D

Zeroes timer

Film commands
= nnn to update parameter
? to query parameter

q

=[8 char string],[?]

q=COPPER

Sets the current film name to COPPER

E

=[0.500 - 99.99],[?]

E=1.23

Set current film density to 1.23

F

=[0.100 - 9.999],[?]

F?

1.23

Returns current film Z factor

J

=[10.0 - 399],[?]

J = 80.1

Set current film tooling to 80.1%

r

=[1 – 50],[?]

r=10

Sets number of samples to 10 for temporal
averaging

Runtime data commands

M

None

M

!

Return crystal fail status, was failed
@ =

crystal good

!

=

crystal failed

S

None

S

-1595

Return thickness value -1.595 kÅ

s

None

s

-1.6

Return film mass -1.6 µg/cm²

T

None

T

12.4

Return rate, 12.4 Å/s

t

None

t

23.4

Return mass accumulation rate in µg*s/cm²,
23.4 µg*s/cm²

U

None

U

5319234.34 Return sensor frequency 5,319,234.34Hz

V

None

V

12.4

Return crystal life, 12.4% remaining

W

None

W

12:45

Return timer, 12 min 45 s

a

None

a

A

Return RESET Status, true
@ =

OK

A

=

lost power

D

=

lost NONV memory

E

=

lost power and NONV memory

NOTE:

NONV refers to non-volatile memory,
i.e. memory that is persistent even
when power is lost. This status is
indicating that there was a loss of
data, either due an internal database
problem, or a database purge.

Table 5-2 STM-2 command set (continued)

STM-2

command

Parameter

Send

Returned

Description