INFICON BPG402-Sx ATM to Ultra-High Vacuum Gauge User Manual

Page 10

Advertising
background image

10

tina46e1-a (2010-03) BPG402.om

Fieldbus name

Profibus

Standard applied

→ [7]

Communication protocol, data format

→ [2], [7]

Interface, physical

RS485

Data rate

≤12 MBaud (→ [2])

Node address

Local
(Adjustable via hexadecimal
"ADDRESS", "MSD", "LSD"
switches)

Default setting

Via Profibus
("ADDRESS" switches set to >7D

hex

(>125

dec

))





00 … 7D

hex

(0 … 125

dec

)

5C

hex



00 … 7D

hex

(0 … 125

dec

)

Profibus connection

D-Sub, 9 pins, female

Cable

shielded, special Profibus cable
(

→ 22 and [5])

Cable length, system wiring

according to Profibus specifications
(

→ [7], [5])

Display panel

Dimensions

Pressure units
Selecting the pressure unit

LCD matrix, 32×16 pixels,
with background illumination

17.0 mm × 12 mm

mbar (default), Torr, Pa
via RS232C (

→ 30)

DANGER

The gauge must only be connected to power supplies, instruments or
control devices that conform to the requirements of a grounded extra-
low voltage (SELV). The connection to the gauge has to be fused
(INFICON-controllers fulfill these requirements).

Supply voltage at the gauge

+24 VDC (+20 … +28 VDC)

1)

ripple max. 2 V

pp

Power consumption

Standard
Degas
Emission start (<200 ms)


≤0.5 A

≤0.8 A

≤1.4 A

Fuse necessary

1.25 AT
((INFICON controllers fulfill these
requirements)

Power consumption

BPG402-S, -SL
BPG402-SD
BPG402-SP


≤18 W

≤18 W

≤20 W

The BPG402-SD requires an additional, separate power supply for the
DeviceNet interface (

→ 21).

Supply voltage at the DeviceNet
connector (Pin 2 and Pin 3)


+24 VDC (+11 … +25 VDC)

Power

consumption

≤2 W

The gauge is protected against reversed polarity of the supply voltage.

Profibus interface
(BPG402-SP)

Display (BPG402-S only)

Power supply