The threshold values of the switching functions are set locally on the potentiome-
ters of the gauge that are accessible via the openings on one side of the gauge
housing.

• Voltmeter
• Ohmmeter or continuity checker
• Screwdriver, max. ø2.5 mm

The procedure for setting thresholds is identical for all switching functions.

Put the gauge into operation.

Connect

the

+ lead of a voltmeter to the threshold measurement point of the

respective switching function (Pin 3 or Pin 6).

Connect the lead of the voltmeter to a ground contact nearby (e.g. locking
screw of the connector, vacuum flange or housing of the gauge).

The threshold voltages are referenced to ground (housing, vacuum
connection), not to Pin 5 (common power GND 24

V supply).

BPG402-S

BPG402-SL

BPG402-SD

BPG402-SP

max. ø2.5

( )

Setpoint Pin 3

max. ø2.5

( )

Setpoint A Pin 3
Setpoint B Pin 6

Using a screwdriver (max. ø2.5 mm), set the voltage of the selected

switching function to the desired value U

Threshold

.

There is no local visual indication of the status of the switching functions.
However, a functional check of the switching functions (On/Off) can be
made with one of the following methods:

• Reading the status via fieldbus interface, for BPG402-SD → [1],

for BPG402-SP

→ [2].

• Measurement of the relay contacts at the sensor cable connector

with an ohmmeter / continuity checker (

→ 19, 20).



4.12.1 Setting the Switching

Functions

Required tools

Procedure