INFICON BPG402-Sx ATM to Ultra-High Vacuum Gauge User Manual

Page 20

Advertising
background image

20

tina46e1-a (2010-03) BPG402.om

TxD

RxD

Degas

42 k

Ω

4

1

11

9

13

14

7

8

2

12

5

15

1.25 AT

24V

Degas

Ident.

RS232

10

-

-

-

Measuring

signal

V

S

6

SP A

SP B

SP A
SP B

3
6

1)

( )

-

Threshold value

8

9

1

15

D-Sub,15 pins,

female,

soldering side

Common (power GND 24V supply)

Ground (housing, vacuum connection)

Electrical connection

Pin 1

Relay switching function A, common contact

Pin 2

Measuring signal output

0 … +10 V

Pin 3

Threshold (setpoint) A

1)

0 … +10 V

Pin 4

Relay switching function A, n.o. contact

Pin 5

Supply common

0 V

Pin 6

Threshold (setpoint) B

1)

0 … +10 V

Pin 7

Degas (active high)

0 V/+24 V

Pin 8

Supply (V

s

) +24

V

Pin 9

Relay switching function B, common contact

Pin 10

Gauge identification

Pin 11

Relay switching function B, n.o. contact

Pin 12

Measuring signal common

Pin 13

RS232C, TxD

Pin 14

RS232C, RxD

Pin 15

Do not connect

1)

Do not connect pin 3 and pin 6 for normal operation of the gauge. These pins

are reserved for adjustment of the setpoint potentiometers (

→ 38).

WARNING

The supply common (Pin 5) and the shielding must be connected at
the supply unit with protective ground.

Incorrect connection, incorrect polarity or inadmissible supply voltages
can damage the gauge.

Sensor cable connection
BPG402-SD, -SP