INFICON BPG402-Sx ATM to Ultra-High Vacuum Gauge User Manual

Page 22

Advertising
background image

22

tina46e1-a (2010-03) BPG402.om

Plug the DeviceNet (and sensor) cable connector into the gauge.

Sensor cable

DeviceNet cable

Lock the DeviceNet (and sensor) cable connector.


For operating BPG402-SP via Profibus, an interface cable conforming to the
Profibus standard is required.

If no such cable is available, make one according to the following indications.

Only a cable that is suited to Profibus operation may be used (

→ [5], [7]).

Make the Profibus interface cable according to the following indications:

1

5

6

9


D-Sub, 9 pins
male, soldering side

Pin 1 Do not connect

Pin 2 Do not connect

Pin

3 RxD/TxD-P

Pin

4 CNTR-P

1)

Pin

5 DGND

2)

Pin

6 VP

2)

Pin 7 Do not connect

Pin

8 RxD/TxD-N

Pin 9 Do not connect

1)

Only to be connected if an optical link module is used.

2)

Only required as line termination for devices at both ends of bus system

(

→ [5]).



3.2.2.3 Making a Profibus

Interface Cable
(BPG402-SP)

Cable type

Procedure