INFICON BPG402-Sx ATM to Ultra-High Vacuum Gauge User Manual
Page 22
22
tina46e1-a (2010-03) BPG402.om
Plug the DeviceNet (and sensor) cable connector into the gauge.
Sensor cable
DeviceNet cable
Lock the DeviceNet (and sensor) cable connector.
For operating BPG402-SP via Profibus, an interface cable conforming to the
Profibus standard is required.
If no such cable is available, make one according to the following indications.
Only a cable that is suited to Profibus operation may be used (
→ [5], [7]).
Make the Profibus interface cable according to the following indications:
1
5
6
9
D-Sub, 9 pins
male, soldering side
Pin 1 Do not connect
Pin 2 Do not connect
Pin
3 RxD/TxD-P
Pin
4 CNTR-P
1)
Pin
5 DGND
2)
Pin
6 VP
2)
Pin 7 Do not connect
Pin
8 RxD/TxD-N
Pin 9 Do not connect
1)
Only to be connected if an optical link module is used.
2)
Only required as line termination for devices at both ends of bus system
(
→ [5]).
3.2.2.3 Making a Profibus
Interface Cable
(BPG402-SP)
Cable type
Procedure