tina46e1-a (2010-03) BPG402.om

23

Plug the Profibus (and sensor) cable connector into the gauge.

Sensor cable

Profibus cable

Lock the Profibus cable (and sensor cable) connector.



The optional 24 VDC power supply (

→ 45) allows RS232C operation of the

BPG402-Sx gauge with any suitable instrument or control device.

The instrument or control device needs to be equipped with a software that sup-
ports the RS232C protocol of the gauge (

→ 30).

Mains connection

Mains voltage

90 … 250 VAC, 50 … 60 Hz

Mains cable

1.8 meter (Schuko DIN and U.S. con-
nectors)

Output (operating voltage of gauge)

Voltage

21 … 27 VDC, set to 24 VDC

Current

Max. 1.5 A

Gauge connection

Connector

D-Sub, 15 pins, female

24 VDC cable

5 m, black

Connection of the instrument or
control device

RS232C connection

D-Sub, 9 pins, female

Cable

5 m, black, 3 conductors, shielded

RS232C
D-Sub, 9 pins

Mains
90 ... 250 VAC
50 ... 60 Hz

13
14

8

15

5

2

3

5

4

6

7

8

N
PE

L

DC

AC

GND

+24 V

PE

BPG402-Sx

D-Sub,

15 pins

3.2.3 Using the Optional

Power Supply
(With RS232C Line)

Technical data

Wiring diagram