2 technical data – INFICON BPG402-Sx ATM to Ultra-High Vacuum Gauge User Manual

Page 8

Advertising
background image

8

tina46e1-a (2010-03) BPG402.om

2 Technical

Data

Measurement range (air, O

2,

CO, N

2

) 5×10

-10

… 1000 mbar, continuous

Accuracy
(after 10 min. stabilization)

15% of reading in the range of
1×10

-8

… 10

-2

mbar

Repeatability
(after 10 min. stabilization)

5% of reading in the range of
1×10

-8

… 10

-2

mbar

Gas type dependence

Appendix B

Switching on threshold
Switching off threshold

2.4×10

-2

mbar

3.2×10

-2

mbar

Emission current

p

≤7.2×10

-6

mbar

7.2×10

-6

mbar < p < 3.2×10

-2

mbar


5 mA
25 µA

Emission current switching

25 µA

⇒ 5 mA

5 mA

⇒ 25 µA


7.2×10

-6

mbar

3.0×10

-5

mbar

Filaments

Number

Means of selection


2

Controlled by gauge (default) or via
interfaces (

→ 33, [1] and [9])

Settling time of measurement signal
after filament change

Filament status


<4 s

LED, relay contact (

→ 28)

Emission control mode

Automatic
Manual


Emission ON/OFF automatically
Emission ON/OFF by user via interfaces
(

→ 29)

Current (p <7.2×10

-6

mbar)

20 mA

Control input signal

0 V/+24 VDC, active high
(

→ 19, 20)

(control via RS232C

→ 30)

Duration

<3 min., followed by automatic stop

In degas mode, the BPG402-Sx gauges keep supplying pressure readings, the
tolerances of which can be higher than during normal operation.

Degas acts only upon the active filament.

Output signal (measuring signal)

0 … +10 VDC

Measuring range

+0.774 … +10 V
(5×10

-10

… 1000 mbar)

Relationship voltage-pressure

logarithmic, 0.75 V/decade
(

Appendix A)

Error signal (

→ 42)

EEPROM error
Hot cathode error
Pirani error

+0.1 VDC

+0.3 VDC

+0.5 VDC

Minimum load impedance

10 k

Ω

BPG402-Sx

42 k

Ω resistor between Pin 10 and Pin 5

(sensor cable)

Measurement

Emission

Degas

Output signal

Gauge identification