INFICON BPG402-Sx ATM to Ultra-High Vacuum Gauge User Manual

Page 36

Advertising
background image

36

tina46e1-a (2010-03) BPG402.om

This interface allows operation of BPG402-SP with part number

353-574 and
353-575

in connection with other devices that are suited for Profibus operation. The physical
interface and communication firmware of BPG402-SP comply with the Profibus
standard (

→ [5], [7].

Two adjustable switching functions are integrated in the BPG402-SP. The corre-
sponding relay contacts are available at the sensor cable connector (

→ 8, 20,

37).

The basic sensor and sensor electronics of all BPG402 gauges are identical.

Caution

Caution: data transmission errors

If the gauge is operated via RS232C interface and Profibus interface at
the same time, data transmission errors may occur.

The gauge must not be operated via RS232C interface and Profibus
interface at the same time.


Via this interface, the following and further data are exchanged in the standardized
Profibus protocol (

→ [2]):

• Pressure

reading

• Pressure unit (Torr, mbar, Pa)
• Degas

function

• Gauge

adjustment

• Status and error messages
• Status of the switching functions

As the DeviceNet protocol is highly complex, the parameters and programming of
BPG402-SP are described in detail in the separate Communication Protocol
(

→ [2]).

For operating the gauge via Profibus, prior installation of the BPG402-SP specific
GSD file is required on the bus master side. This file can be downloaded via inter-
net (

→ [3]).

For unambiguous identification of the gauge in a Profibus environment, a node ad-
dress is required. The node address setting is made on the gauge.

The node address (0 … 125

dec

) is set in hexadecimal form

(00 … 7D

hex

) via the "ADDRESS", "MSD", and "LSD" switches.

The node address is polled by the firmware when the gauge is
switched on. If the setting deviates from the stored value, the
new value is taken over into the NVRAM. If a value >7D

hex

(>125

dec

) is entered, the node address setting currently stored in

the device remains valid but it can now be defined via Profibus
("Set slave Address",

→ [2]).

Default address setting is 5C

hex

.

The gauge is connected to Profibus via the 9-pin Profibus connector (

→ 22).



4.11 Profibus

Interface

(BPG402-SP)

4.11.1 Description of the

Functions

4.11.2 Operating

Parameters

4.11.2.1 Operating

Software

4.11.2.2 Node Address Setting

Electrical connections