tina46e1-a (2010-03) BPG402.om

31

• TxD

Pin 13

• RxD

Pin 14

• GND

Pin 5

(Sensor cable connector)

The complete output string (frame) is nine bytes (byte 0 … 8). The data string is
seven bytes (byte 1 … 7).

Byte No Function

Value

Comment

0

Length of data string

7

set value

1

Page number

5

hot cathode gauges

2 Status

→ Status byte

3 Error

→ Error byte

4

Measurement high byte

0 … 255

→ Calculation of pressure value

5

Measurement low byte

0 … 255

→ Calculation of pressure value

6

Software version

0 … 255

→ Software version

7

Sensor type

12

for BPG402-Sx

8

Check sum

0 … 255

→ Synchronization

Synchronization of the master is achieved by testing three bytes:

Byte No Function

Value

Comment

0

Length of data string

7

set value

1

Page number

5

hot cathode gauges

8

Check sum of bytes No 1 … 7

0 … 255

low byte of
check sum

1)

1)

High order bytes are ignored in the check sum.

Bit 1

Bit 0

Definition

0 0

emission

off

0 1

emission 25

μA

1

0

emission 5 mA

1 1

degas

Bit 2

Definition

x

not

used

Bit 3

Definition

0

⇔ 1

toggle bit, changes with every
string received correctly

Bit 5

Bit 4

Definition

0

0

current pressure unit mbar

0

1

current pressure unit Torr

1

0

current pressure unit Pa

Bit 6

Definition

0

filament 1 active

1

filament 2 active

Bit 7

Definition

x

not

used




Electrical connections

4.9.1.1 Output String (Transmit

)

Format of the output string

Synchronization

Status byte